Metrohm 930 Compact IC Flex Oven/SeS/Deg User Manual

Page 58

Advertising
background image

3.20 Connecting the separation column

■■■■■■■■■■■■■■■■■■■■■■

50

■■■■■■■■

930 Compact IC Flex Oven/SeS/Deg (2.930.2460)

Connecting the separation column

1 Removing the stoppers

CAUTION

When inserting the column, ensure that it is inserted correctly
based on the marked flow direction.

■

Remove the stoppers from the separation column.

2 Installing the input of the separation column

There are 3 possibilities:

■

Attaching the bottom end of the separation column directly to
the guard column.
or

■

If the guard column is connected to the separation column using
a provided connection capillary, connect the bottom end of the
separation column to the guard column's output capillary using
the provided PEEK pressure screw (6.2744.070).
or

■

If no guard column is used (not recommended), connect the col-
umn input capillary to the input of the separation column using a
short pressure screw (6.2744.070).

3 Rinsing the separation column

■

Place a beaker under the separation column's outlet.

■

Start manual control in MagIC Net and select the high-pressure
pump: Manual

▶ Manual control ▶ Pump

– Flow: in accordance with column leaflet
– On

■

Rinse the separation column with eluent for approximately 10
minutes.

■

Use manual control in MagIC Net to stop the high-pressure pump
again: Off.

4 Removing the coupling

■

Remove the coupling (6.2744.040) from the column output capil-
lary.