2.3 Setting the DC output voltage

Relay Box 731, Instructions for use

9

2.3 Setting the DC output voltage

The common output voltage for the direct current outputs (DC1 and
DC2) is set inside the Relay Box housing.

•

Switch off the instrument and pull out the line plug. Loosen the
four knurled screws of the protective cover and remove them.

Right-hand side
of housing with-
out protective
cover

Rear view of
housing without
protective cover

•

The screws marked with an arrow in the diagram must be loos-
ened.

•

Carefully remove the instrument PCB from the housing.

To set the output voltage the plug-in bridge P9 on the relay box
motherboard must be fitted with the appropriate jumper .

The location of plug-in bridge P9 can be seen in the diagram of the
instrument motherboard shown below.

+18V

+10V

+24V

0V*
+5V

1

12

P9

6

7

Set the jumper, which is located in the +24 V position
as standard, to the position corresponding to the re-
quired output voltage with the help of tweezers or ta-
per-nose pliers.

+18V

+10V

+24V

0V*

+5V

1

12

P9

6

7

Carefully reinsert the motherboard in the housing and screw the in-
strument together. Use the adhesive labels provided to note the set
output voltage on the rear panel of the instrument.

Jumper