2.4 Connections
Relay Box 731, Instructions for use
17
Settings for the 731 Relay Box:
772 Pump Unit connected to DC1 (or DC2)
Output voltage:
18 or 24 volt
Remote address selection disk DC1:
7 (=Output 7)
Thermostatic bath connected to AC1 (or AC2, line supply)
Remote address selection disk AC1:
6 (=Output 6)
711 Liquino, control commands:
SEQ mode
Switch on pump:
Command: CONTROL
Interface: Remote
Signals 1***----
(Output 7)
(switch off with 0***----)
Switch on thermostatic bath:
Command: CONTROL
Interface: Remote
Signals *1**----
(Output 6)
(switch off with *0**----)
The output lead can be selected freely. Avoid conflicts with other in-
struments which may be connected.
751 Titrino with 700 Dosino – 731 Relay Box –
772 Pump Unit
The 772 Pump Unit can be used aspirate the contents of a titration
vessel.
731
772
751
Settings for the 731 Relay Box:
772 Pump Unit connected to DC1 (or DC2)
Output voltage:
18 or 24 volt
Remote address selection disk DC1:
4 (=Output 4)
6.2125.100 cable