Metrohm 731 Relay Box User Manual
Page 22

2. Installation
Relay Box 731, Instructions for use
16
726 Titroprocessor – 731 Relay Box –
2x hotplates/stirrers
Control of hotplates/stirrers for titrations with the 726 Titroprocessor.
731
726
Settings for the 731 Relay Box:
Connected to AC1 and AC2
(line voltage)
Remote address selection disk AC1:
6 (=Output 6)
Remote address selection disk AC2:
7 (=Output 7)
726 Titroprocessor, control commands:
Switching on the hotplates/stirrers:
CTRL_RM Remote A Signals 1*******
CTRL_RM Remote A Signals *1******
Switching off both hotplates/stirrers:
CTRL_RM Remote A Signals 00******
The output lead can be selected freely. Avoid conflicts with other in-
struments which may be connected.
711 Liquino – 731 Relay Box – 772 Pump Unit –
Thermostat
Liquid handling setup with 772 Pump Unit as delivery or aspirating
pump and a thermostatic bath.
711
731
772
6.2125.100 cable
6.2125.100 cable
Dosinos
Dosinos
Thermostatic bath