Metrohm 731 Relay Box User Manual

Page 18

Advertising
background image

2. Installation

Relay Box 731, Instructions for use

12

Assignation of the remote socket according to pins:

Pin

I/O lead

Function

1

Output 6

L3 activate

2

Output 7

Pulse for recorder

3

Output 4

L1

4

Output 2

Determination busy

5

Output 0

Ready

6

Output 8

not used

7

Output 9

not used

8

Output 10

not used

9

Input 1

Stop

10

Input 3

Clear

11

Input 5

not used

12

Input 7

not used

13

Output 11

not used

14

0 Volt

15

+5 Volt

16

Output 5

Error

17

Output 3

EOD (End of Determination)

18

Output 1

Conditioning ok

19

Output 12

not used

20

Output 13

not used

21

Input 0

Start

22

Input 2

Enter

23

Input 4

Sample Ready

24

Input 6

not used

25

0 Volt

The pin assignation or the allocation of functions of the remote lines
may differ slightly from one Metrohm instrument to another. Consult
the ‘Instructions for use’ of your instrument.