Figure 8, Mounting the air tube: correct - incorrect, Figure 9 – Metrohm 893 Professional Biodiesel Rancimat User Manual

Page 25: Mounting the foam barrier

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

893 Professional Biodiesel Rancimat

■■■■■■■■

17

Figure 8

Mounting the air tube: correct - incorrect

WARNING

The foam barrier can melt if it projects too deeply into the heating
block.

Ensure that the foam barrier is at least 7 cm above the base of the
reaction vessel.

Figure 9

Mounting the foam barrier

3.2.4

Inserting vessels / Establishing tubing connections

After you have assembled the reaction and measuring vessels, insert them
in the 893 Professional Biodiesel Rancimat and establish the tubing con-
nections (see Chapter 2.1, page 6) as follows:

1 Inserting the measuring vessel

■

Fill distilled water into the measuring vessel.

■

Place the measuring vessel cover onto the measuring vessel.

■

Insert the measuring vessel into the openings provided on the 893
Professional Biodiesel Rancimat. While doing so, carefully guide
the connector plug into the electrode connector.

■

Connect the black Iso-Versinic

®

tubing to the M8/olive tubing

adapter of the measuring vessel cover.