4 cleaning the oil separator, Cleaning the oil separator – Metrohm 893 Professional Biodiesel Rancimat User Manual

Page 40

Advertising
background image

4.3 Determinations

■■■■■■■■■■■■■■■■■■■■■■

32

■■■■■■■■

893 Professional Biodiesel Rancimat

■

Clean the measuring vessel covers, the PTFE cannula and the
electrodes with acetone or 2-propanol or in the dishwasher and
thoroughly rinse with distilled water.
Pre-clean with dishwashing detergent in the case of severe con-
tamination.
Remove the protective ring to facilitate cleaning of the electrodes.

2 Cleaning the reaction vessels and accessories

■

Dispose of used reaction vessels and air tubes and use new
reaction vessels and air tubes for the next measurement.

■

Clean the reaction vessel covers with acetone or 2-propanol or
in the dishwasher and rinse with distilled water.
Pre-clean with dishwashing detergent in the case of severe con-
tamination.

■

Then heat the reaction vessel covers in the drying oven for 2
hours at 80 °C.

NOTE

Replace the reaction vessel covers if they no longer sit tightly on
the reaction vessel or the material is brittle.

3 Cleaning the tubing

■

Clean the Iso-Versinic

®

tubing with acetone or 2-propanol or in

the dishwasher and rinse with distilled water.
Pre-clean with dishwashing detergent in the case of severe con-
tamination.

■

Then heat the Iso-Versinic

®

tubing in the drying oven for 2

hours at 80 °C.

4.3.4

Cleaning the oil separator

The oil separator is available as an optional accessory.