background image

■■■■■■■■■■■■■■■■■■■■■■

4 Operation

893 Professional Biodiesel Rancimat

■■■■■■■■

31

4 Connecting and inserting reaction vessels

NOTE

The temperature defined in the method has to be reached before
you insert the reaction vessels, i.e., the temperature display has to
be green.

■

Seal the unused positions with stoppers or empty reaction vessels
for protection against impurities.

■

Connect the Iso-Versinic

®

tubing which is connected to the reac-

tion vessel covers to the M8 tubing adapter/olive of the measur-
ing vessel covers.

■

Screw the pieces of FEP tubing 250 mm to the M8/M6 thread
adapters of the reaction vessel covers and the air supply connec-
tors of the 893 Professional Biodiesel Rancimat.

■

Insert the prepared reaction vessels into the recesses of the heat-
ing block.

■

Press the start button to start the data recording immediately
after inserting each single reaction vessel.

4.3.3

Cleaning the accessories

1 Cleaning the measuring vessels and accessories

■

Pre-clean with dishwashing detergent in the case of severe con-
tamination.

■

Clean used measuring vessels after pouring off the measuring
solution with acetone or 2-propanol or in the dishwasher.
Pre-clean with dishwashing detergent in the case of severe con-
tamination.

■

Thoroughly rinse with distilled water.

■

Clean the measuring vessel covers, the PTFE cannula and the
electrodes with acetone or 2-propanol or in the dishwasher and
thoroughly rinse with distilled water.
Pre-clean with dishwashing detergent in the case of severe con-
tamination.
Remove the protective ring to facilitate cleaning of the electrodes.

■

Clean used measuring vessels after pouring off the measuring
solution with acetone or 2-propanol or in the dishwasher.
Pre-clean with dishwashing detergent in the case of severe con-
tamination.

■

Thoroughly rinse with distilled water.