3.2 Mounting accessories

■■■■■■■■■■■■■■■■■■■■■■

18

■■■■■■■■

893 Professional Biodiesel Rancimat

2 Mounting the tubing for the air supply

■

Screw the FEP tubing 250 mm to the air supply connections of
the 893 Professional Biodiesel Rancimat.

3 Inserting the reaction vessel

■

Fill the reaction vessel with the sample.

■

After the required reaction temperature has been reached, insert
the reaction vessel with the mounted reaction vessel cover in the
openings provided on the 893 Professional Biodiesel Rancimat.

4 Establishing the tubing connections

■

Connect the Iso-Versinic

®

tubing that is connected to the measur-

ing vessel cover to the tubing connector of the reaction vessel
cover.

■

Screw the FEP tubing 250 mm which is connected to the M8/olive
tubing adapter of the 893 Professional Biodiesel Rancimat to the
M8/M6 thread adapter of the reaction vessel cover.

3.2.5

Mounting the exhaust air collection tube

The optional exhaust air collection tube (6.2757.000)

[Link target not

found in publication context!]

can be mounted on the 893 Professional

Biodiesel Rancimat for targeted removal of the exhaust air.

NOTE

A total of 8 pieces of Iso-Versinic

®

tubing (6.1839.000) are required in

addition to the exhaust air collection tube.

Proceed as follows to mount the collection tube:

1 Mounting the exhaust air collection tube

■

Insert the exhaust air collection tube with both support pins into
the collection tube holders on the 893 Professional Biodiesel Ran-
cimat in such a way that the connector to the exhaust air removal
is located to the rear.

2 Connecting the measuring vessels

■

Screw the M8/olive tubing adapter into the Out opening of the
measuring vessel cover.

■

Connect one end of the Iso-Versinic

®

tubing to the M8/olive tub-

ing adapter.