4 housing, 5 leak sensor, 6 high-pressure pump – Metrohm 930 Compact IC Flex SeS User Manual

Page 100: Housing, Leak sensor, High-pressure pump

Advertising
background image

7.4 Housing

■■■■■■■■■■■■■■■■■■■■■■

92

■■■■■■■■

930 Compact IC Flex SeS (2.930.1400)

7.4

Housing

Dimensions

Width

302 mm

Height

562 mm

Depth

368 mm

Material of base
tray, housing and
bottle holder

Polyurethane hard foam (PUR) with flame retardation for fire class
UL94V0, CFC-free, coated

Operating ele-
ments

Indicators

LED standby indicator

On/off switch

On the rear of the instrument

7.5

Leak sensor

Type

Electronic, no calibration necessary

7.6

High-pressure pump

Type

Serial dual-piston pump

Intelligent pump head recognition

Chemically inert

Metal-free pump heads

Materials in contact with the eluent: PEEK, ZrO

2

, PTFE/PE

Self-optimizing flow and pressure

Flow rate

Adjustable flow
range

0.001 - 20.0 mL/min

Flow increment

1 µL/min

Reproducibility
of the eluent
flow

< 0.1% deviation

Pressure range

Pump

0 - 50.0 MPa (0 - 500 bar)

Pump head

0 - 35.0 MPa (0 - 350 bar) (applies for the standard PEEK pump head)

Residual pulsa-
tion

< 1%

Advertising