Figure 8, Installing tubing weighting and aspiration filter – Metrohm 930 Compact IC Flex SeS User Manual

Page 29

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

930 Compact IC Flex SeS (2.930.1400)

■■■■■■■■

21

■

Place the loose end of the eluent aspiration tubing into the aspira-
tion filter.
The end of the tubing should reach approximately to the center of
the aspiration filter.

■

Tighten the aspiration filter to the filter holder.

Figure 8

Installing tubing weighting and aspiration filter

3 Installing the eluent bottle cap on the eluent bottle

■

Insert the eluent aspiration tubing into the eluent bottle
(6.1608.070).

■

Tighten the bottle cap on the eluent bottle.

■

Adjust the length of the eluent aspiration tubing so that the aspi-
ration filter is at the bottom of the eluent bottle. Then fasten it in
place using the M8 tubing nipple.

■

Seal the M6 opening on the bottle cap with the M6 threaded
stopper from the accessory set.