3.5 Connecting the eluent bottle

■■■■■■■■■■■■■■■■■■■■■■

20

■■■■■■■■

930 Compact IC Flex SeS (2.930.1400)

■

The eluent bottle cap GL 45 accessory set (6.1602.160)
This accessory set contains the bottle cap, an M6 tubing nipple, an M8
tubing nipple, two O-rings and an M6 and M8 threaded stopper.

■

The tubing adapter for aspiration filter accessory set (6.2744.210)
This accessory set contains a filter holder, a clamping screw and tubing
weighting.

■

An aspiration filter (6.2821.090)

Connecting the eluent aspiration tubing

1 Installing the eluent bottle cap (6.1602.160)

■

Start by pushing the M8 tubing nipple onto the loose end of the
eluent aspiration tubing, followed by the O-ring.

■

Push the loose end of the eluent aspiration tubing through the
M8 opening of the bottle cap and screw it on for the time being.

Figure 7

Installing the eluent bottle cap

2 Mounting the tubing adapter and the aspiration filter

Install the parts of the tubing adapter for aspiration filter
(6.2744.210) accessory set:

■

Start by pushing the tubing weighting onto the loose end of the
eluent aspiration tubing.

■

Then push the clamping screw onto the loose end of the eluent
aspiration tubing.

■

Lastly, push the filter holder onto the loose end of the eluent aspi-
ration tubing and screw it onto the tubing nipple.
The end of the tubing should extend approximately 1 cm.

Installing the aspiration filter:

NOTE

Only handle the aspiration filter while wearing gloves.

Advertising