Problem
Cause
Remedy
Guard column – Dimin-
ished efficiency.
■
Replace the guard column (see Chapter
3.16, page 45).
Separation column – Dead
volume at the column
head.
■
Install the separation column in the oppo-
site flow direction (if permitted by the leaf-
let) and rinse into a beaker.
■
Replace the separation column (see "Con-
necting the separation column", page 49).
Precision problems -
the measured values
are highly scattered.
Injection valve – Sample
loop.
Inspect the installation of the sample loop (see
"Optional: Exchanging the sample loop",
page 26).
Sample – There are gas
bubbles in the sample.
Use the sample degasser .
Sample – The rinsing vol-
ume is too small.
Increase the rinsing time (see Chapter 5.13,
page 83).
Injection valve – Defective.
Request Metrohm Service.
MCS – The vacuum is too
low.
■
Check the connectors. If they are OK:
■
Contact Metrohm Service.