Metrohm 788 IC Filtration Sample Processor User Manual

Page 142

Advertising
background image

6 Appendix

788 IC Filtration Sample Processor

134

Error message

Cause

* changer low power

The power supply cannot deliver

enough power for the simultaneous

operation of all components currently

in use (pump, lift).

* RS232 error

The transmission parameters of the

RS232 interface do not agree with

those of the receiving instrument.

* check rack/swing head

Invalid configuration of sample rack

or swing head.

* invalid position

The sample position selected is de-

fined as a special beaker or the spe-

cial beaker chosen is not defined.

* invalid rack code

The rack code read by the 788 IC Fil-

tration Sample Processor could not

be found in the internal position ta-

bles.

* changer not ready

The 788 IC Filtration Sample Proces-

sor cannot execute the command

chosen because it is busy carrying

out another action or it cannot move

to the rack position.

* changer overload

Load or resistance too large to carry

out the chosen action.

trap error xxx

Unexpected program error, turn in-

strument off and on again.

No display, LEDs tower

1 and tower 2 are lit up

LCD error (system error 7). Please in-

form Metrohm Service.