5 ic system with sample dialysis – Metrohm 788 IC Filtration Sample Processor User Manual
Page 48
2 Installation
788 IC Filtration Sample Processor
40
2.6.5
IC system with sample dialysis
The 788 IC Filtration Sample Processor can be connected to an IC
system with sample dialysis. However, this application is not planned
for usage together with the ultra-filtration cell, which should be bridged.
The sample processor is connected to the modular IC system
consisting of 732 IC Detector, 733 IC Separation Center, 709 IC Pump,
754 Dialysis Unit and (if suppression is used) the 753 Suppressor
Module as shown in Fig. 22 using the 6.2125.120 adaptor. With this
interconnection the standard method
'Dialysis'
can be used (see
section 4.6). If no suppression is used, the 754 Dialysis Unit can be
connected to the 6.2125.120 adaptor using the 6.2143.200 cable.
732
733
709
753
754
788
Abb. 22: Interconnection with IC system with dialysis
6.2125.090 cable
6.2125.060 cable
6.2125.090 cable
6.2125.120 adaptor
6.2143.220 cable
6.2141.110 cable