788 IC Filtration Sample Processor

23

2.4

Installation for full control with «IC Net»

2.4.1 Connection

options

The 788 IC Filtration Sample Processor can be connected to Metrohm

IC Systems according to the following techniques:

1. The IC System is totally controlled by the PC software

Data acquisition and direct control of the IC components is done by the

PC software «IC Net 2.1» in combination with the 762 IC Interface. The

788 IC Filtration Sample Processor is connected to the IC system both

by the RS232 Interface and by the remote interface. Due to direct

control of all procedures by the PC software, internal methods of the

788 IC Filtration Sample Processor are not used. This section describes

several example combinations of the 788 IC Filtration Sample

Processor with Metrohm IC systems based on this technique.

2. Compact IC Systems are remote controlled by PC software

Compact designed IC-Systems 761 Compact IC and 790 Personal IC

are also controlled by the PC. The 788 IC Filtration Sample Processor is

connected to these IC Systems by remote connection only. Therefore,

internal methods of the sample processor are used. Please find a

description of the connection of the 788 IC Filtration Sample Processor

to the IC Systems 761 Compact IC and 790 Personal IC in section 2.5.

The required 788 standard method is given in section 4.6.3.

3. The IC System is connected to PC for simple control

Instead of using the 762 IC Interface, modular Metrohm IC Systems can

be simply controlled via the 714 PC Board or 771 IC Compact Interface.

These devices just convert the analog signals from the 732 IC Detector

to digital data for the PC. For simple control of the IC system, the 732 IC

detector has to be connected to the PC additionally via its RS232 port.

«IC Net» and the prior software «IC Metrodata» use this serial

connection for a simple control of the IC system. As described for

compact IC systems the 788 IC Filtration Sample Processor is

connected to these IC Systems by remote connection. Also, internal

methods of the Sample Processor are used. Please find a description

of the connection of the 788 IC Filtration Sample Processor to such IC

Systems in section 2.6. The use of the corresponding 788 standard

methods is given in section 4.6.