4 Operation

788 IC Filtration Sample Processor

86

4.3.8

User defined methods

USER

METHOD

The <USER METHOD> key is used for the management of user de-

fined methods, which can be stored and recalled under freely select-

able names. A method comprises the parameters defined under the

<PARAM> key. The key opens the following main menu:

methods
>recall method

Recall method

methods
>store method

Store method

methods
>delete method

Delete method

Repeated pressing of the <USER METHOD> key selects the sub-

menus in turn. Use <Ð> or <Ï> to move up or down one menu item;

use <HOME> or <END> to move to the first or last menu item re-

spectively. The submenu is opened with <ENTER>, <QUIT> returns

to the normal state. The following listings show all dialog items which

appear under <USER METHOD>. The values shown in the display are

the default values, the possible entry values or ranges are shown below

the display.

methods
>recall method

Recall existing method

>recall method
method: ********

8 ASCII characters

Selection of the method

With <SELECT> any method saved can be chosen. If

an "empty" method is to be loaded, the method

'********'

can be selected. Like this, the actual work

memory is deleted.


methods
>store methods

Storage of methods

>store methods
method: ********

8 ASCII characters

Define method name

"<" or ">" activates text edit mode where any method

name desired can be entered (see section 4.1.6).
When a certain method should be worked off after

switching on the 788 IC Filtration Sample Processor, a

command sequence can be stored under the name

'POWERUP'

. This method is started automatically after

switching on the main switch (see section 4.3.9).

Advertising