Metrohm 714 IC Metrodata User Manual

Page 29

Advertising
background image

3.3 Use of existing methods

IC Metrodata for Win95

23

Preparations

•

Insert separating column in IC system and condition it.

•

Prepare standards and sample.

Load standard method and run

•

Click on

or select

Load method and run

from the

Acquire

menu.

•

Select the desired method (e.g.

dual1sup.mtw

) in the

File Open

win-

dow and click the

OK

button. The window

Start Run

appears.

•

Enter a name as

Ident

to characterize the standard (max. 14 char-

acters, e.g. "standard").

•

Change the analysis time

Duration

if required.

•

Enter more parameters to characterize sample and acquirement

conditions if required.

•

Click the

OK

button.

The window for the on-line display of the chromatogram opens.

The window title line contains sample name, method and run num-

ber (e.g.

standard (dual1sup.mtw)01.run

). The status display below

shows the message

Waiting

.

•

Set injection valve at 733 IC Separation Center to "FILL" (manually

or by program) and fill sample loop with standard.

•

Set injection valve at 733 IC Separation Center to "INJECT" (manu-

ally or by program). The data acquisition is started automatically

and stopped as soon as the analysis time has passed. As the

Re-

start method

option is switched on in the method, the chroma-

togram is saved automatically and deleted in the chromatogram

window.