Metrohm 714 IC Metrodata User Manual

Page 54

Advertising
background image

3 Learning by doing

IC Metrodata for Win95

48

•

Select the IC system type in the

Initial

tab. With

Anion

and

Cation

temperature coefficient and polarity are set automatically, with

User

these values can be combined independently.

•

Change the parameters for 732 IC Detector, 733 IC Separation

Center and 709 IC Pump to the values to be set for these instru-

ments at the start of the method.

•

First open the

Program

tab, then open the

Text

tab. Click the

Add

button to add a new program line.

•

Enter the time

x.x

in min into the first field of the program line.

•

Select the desired program point in the second field.

•

Select the parameter belonging to the program point in the third

field. For

Flow

,

Pmax

and

Pmin

values must be entered, for

Remote

1

,

0

or

*

must be entered. A certain remote line position can be

changed by moving the cursor in front of this position and then en-

tering

1

,

0

or

*

.

•

Click the

Add

button to add new program lines or click the

Delete

button to delete program lines. Faulty program points are dis-

played in the lower part of the program field.

•

If you wish to use predefined remote instructions, open the

Remote

configuration

tab.

•

Click the

Add

button to add a new remote instruction.

•

Enter any name you like for the remote instruction (max. 19 charac-

ters, e.g.

Ext. Valve Fill

) in the first field.

•

Enter

1

,

0

or

*

for the corresponding remote instruction for remote

control lines 1...8 in the second field. A certain remote line position

can be changed by moving the cursor in front of this position and

then entering

1

,

0

or

*

.

•

Click the

Add

button to add more remote instructions or click the

Delete

button to delete remote instructions.

•

Click the

Activate

button. The entered remote instruction are added

to the standard program points and can also be selected.

•

Open the

Text

tab and add the predefined remote instructions into

the time program.

•

After finishing the programming, click the

Apply

button. The pro-

gram and the parameters are tested for consistency.

•

Switch on the

ENABLED

option and close the window with

OK

.

•

Close the

Equipment

window with

Close

.

•

Select

Save method as

of the

File

menu and save the changed

method under the same or a new name.