3.4 Multipoint calibration

IC Metrodata for Win95

35

Save method

•

Select

Save method as

of the

File

menu. The

File Save As

window

appears.

•

Click the

OK

button. A window appears with the message

c:\..... al-

ready exists. Overwrite?

•

Click the

OK

button.

Record a sample chromatogram

•

Click on

or select

Restart method

of the

Acquire

menu.

•

Enter a sample name (max. 14 characters, e.g. "drinking water") as

Ident

in the

Start Run

window.

•

If required, enter more parameters (e.g.

Dilution

) to characterize

sample and acquirement conditions.

•

Click the

OK

button.

The window for the on-line display of the chromatogram opens.

The status display below shows the message

Waiting

.

•

Set the injection valve at the 733 IC Separation Center to "FILL"

(manually or by program) and fill the sample loop with sample.

•

Set the injection valve at the 733 IC Separation Center to "INJECT"

(manually or by program). The data acquisition is started automati-

cally and stopped as the analysis time has passed.