3.5 Modify several chromatograms

IC Metrodata for Win95

37

•

Select the desired method for recalculation of all chromatograms

recorded with this method in the

Chromatograms acquired by method

field.

•

If the sample chromatograms (Level = 0) should be recalculated

switch on the

Sample runs

option.

•

If the standard chromatograms (Level

≥

1) should be recalculated

switch on the

Calibration runs

option.

•

If the chromatograms should be reintegrated switch on the

Reinte-

grate

option (if the queue contains standards, the

Recalibrate

option

is switched on automatically). When the reprocessing is started the

Integration parameters

window appears (procedure see section 3.2

– Modify integration parameters).

•

If a new calibration should be performed switch on the

Recalibrate

option. In this case, select the desired options in the

Recalibration

part of the window.

•

If a report should be printed out automatically for all recalculated

chromatograms switch on the

Print report

option. When the reproc-

essing is started the

Report options

window appears (procedure

see section 3.2 – Change report options).

•

Click the

Start

button.