Metrohm 714 IC Metrodata User Manual

Page 62

Advertising
background image

3 Bedienungslehrgang

IC Metrodata for Win95

56

•

Select

Equipment setup

of the

Options

menu.

•

In the

Configured equipment control drivers

window, click on

732 IC

Detector[ COM# ]

and afterwards on

Setup

. The

INITIAL

tab in the

732

IC Detector[ COM# ]

window is opened.

•

Open the

Manual Control

tab.

•

Program RUN/STOP

: The program at the 732 IC Detector is

started or stopped.

•

Valve A FILL/INJECT

: Valve A at the 733 IC Separation Center is

switched to the

Fill

or

Inject

position.

•

Valve B FILL/INJECT

: Valve B at the 733 IC Separation Center is

switched to the

Fill

or

Inject

position (only

possible with the 733.0020 IC Separation

Center).

•

Suppressor STEP

:

The suppressor is switched to the next

position (only possible with the 733.0030

IC Separation Center).

•

Zero ON/OFF

:

The auto-zero is switched on or off.

•

Pump RUN/STOP

:

The 709 IC Pump is started or stopped.

Log data transmission
The data transmission logging is used for error diagnosis if the com-

munication between PC and 732 IC Detector is faulty. Normally, it

should not be used because the data transmission rate is lowered.

•

Select

Equipment setup

of the

Options

menu.

•

In the

Configured equipment control drivers

window, click on

732 IC

Detector[ COM# ]

and afterwards on

Setup

. The

INITIAL

tab in the

732

IC Detector[ COM# ]

window is opened.

•

Open the

Serial port[ COM# ]

tab.

•

Select the log type:

Formatted I/O

: The data transmission between PC and 732 IC

Detector 732 is logged line by line.

Byte stream

: The data transmission between PC and 732 IC

Detector 732 is logged character by character

(1 character per line).

File

:

The data transmission logging is saved into the

MemDump.log

file in the program directory.