2 Installation

753 Suppressor Module / 8.753.1003 Instructions for Use

12

7 Attach pump tubings

•

Loosen both tubing cartridges 44 from the holding clamp 99 by

pressing down snap-action lever 10

10 and remove from

mounting pin 12

12 on the 753 Suppressor Module (see Fig. 1).

•

Press contact pressure lever 55 on both tubing cartridges

down as far as it will go.

•

Insert a length of pump tubing 33

33 (6.1826.050) into each of

the tubing cartridges as shown in Fig. 7. The white-yellow

stopper 34

34 must click into the corresponding holder on the

left-hand side of the tubing cartridge.

•

Place the tubing cartridges on mounting pin 12

12 and press

down on the right-hand side until snap-action lever 10

10 clicks

into position on holding clamp 99. Take care that no kinks are

formed in the pump tubing.

8 Suppressor connection 2: H

2

SO

4

•

Pull the inlet capillary 25

25 marked with “H

2

SO

4

“ at suppressor

connection 22

22 (see Fig. 6) through one of the rear panel

openings 29

29 of the 690 Ion Chromatograph or through the

door gap in the right-hand lower corner of the interior door

13

13.

•

Mount a PEEK compression fitting 31

31 (6.2744.010) on the

end of inlet capillary 25

25 and screw this on to coupling 32

32

(6.2744.030) (see Fig. 7).

•

Push coupling 32

32 on to the outlet end of the first pump tubing

33

33 (see Fig. 7).

•

Mount suction tubing: cut a piece of PTFE Tubing 30

30

(6.1803.020) to the desired length.

•

Mount a PEEK compression fitting 31

31 (6.2744.010) on one

end of PTFE tubing 30

30 and screw this on to coupling 32

32 (see

Fig. 7).

•

Push coupling 32

32 on to the inlet end of the first pump tubing

33

33 (see Fig. 7).

•

Immerse the other end of the suction tubing in a vessel

containing regeneration solution (normally 20 mmol/L H

2

SO

4

)

and fix in place.

•

Pull outlet capillary 26

26 marked with "Waste" at suppressor

connection 22

22 (see Fig. 6) through one of the rear panel

openings 29

29 of the 690 Ion Chromatograph.

•

Lead outlet capillary 26

26 to a sufficiently large waste container

and fix in place.