Metrohm 753 Suppressor Module User Manual
Page 34

4.2 Maintenance and servicing
753 Suppressor Module / 8.753.1003 Instructions for Use
31
38
38
37
37
36
36
35
35
Fig. 9: Assembling the suppressor
35
35
Screw nut
37
37
Suppressor rotor (6.2832.000)
36
36
Connection piece
(6.2832.010)
with input and output leads
38
38
Suppressor holder
6 Clean connection piece
•
Clean the sealing surface of connection piece 36
36 with the aid
of a lint-free cloth and ethanol.
7 Insert connection piece
•
Insert connection piece 36
36 in suppressor holder 38
38 in such a
way that connection "1" is at the top and that the three lugs on
the connection piece fit in the corresponding openings of the
holder.
•
Screw nut 35
35 onto the thread of suppressor holder 38
38 manu-
ally (do not use tools).
8 Connect and condition the suppressor
•
Reconnect the suppressor to the IC system (see section 2.3.2
and section 2.4.2).
•
Before switching the suppressor to the next position for the
first time rinse all 3 suppressor units with solution for 5 min.