2 Installation

753 Suppressor Module / 8.753.1003 Instructions for Use

14

•

Pull outlet capillary 27

27 marked with "Waste" at suppressor

connection 22

22 (see Fig. 6) through one of the rear panel

openings 29

29 of the 690 Ion Chromatograph.

•

Lead outlet capillary 27

27 to a sufficiently large waste container

and fix in place.

10 Startup of 753 Suppressor Module

•

Set switch <REMOTE> 22 on 753 Suppressor Module to

"OFF".

•

Switch on 753 Suppressor Module with mains switch 13

13 and

set switch <PUMP> 33 to "ON".

•

Adjust the contact pressure for both tubing cartridges: press

contact pressure lever 55 upwards until the solutions just start

to be drawn in. Then press the contact pressure lever up-

wards until it clicks once more to obtain optimal contact pres-

sure.

•

Before switching the suppressor to the next position for the

first time rinse all three suppressor units with solution for

5 min.

•

Check all tubing from the storage vessels through the tubing

cartridges and the suppressor up to the waste containers for

leakage of liquids. If liquid escapes anywhere the corre-

sponding compression fitting must be tightened further or

changed.

Pump tubings are consumable material with a lifetime which depends

on the contact pressure. This is why the tubing cartridges should be

raised completely by loosening snap-action lever 10

10 on the right-hand

side if the pump is to remain switched off for a considerable length of

time (the set contact pressure remains unchanged).