9 cal calibration of ph sensors – Metrohm 796 Titroprocessor User Manual

Page 123

Advertising
background image

7. Titration and measuring modes

796 Titroprocessor, Instructions for use

117

117

7.9 CAL

Calibration of pH sensors

Application:

The calibration of a pH sensor is indispensable not only for pH measure-
ments; it is also essential for

SET titrations and for the correct determination

of pK values and fixed endpoints.
In contrast to manual sensor calibration, the

CAL mode allows the calibra-

tion parameters and individual buffer series to be set down in a method. The
calibration data, like all determination data, can be stored and are available
for subsequent data output; however, they cannot be reactivated.

Calibration principle

During the pH calibration an isotherm (constant temperature) of the pH
value is recorded as a function of the electrode system voltage (electrode
system = sensor + electrode cable + connection socket + measurement
interface) and used for the conversion of the measured voltages into pH
values. This means that all the buffer solutions used for the calibration must
be at the same temperature, even with automatic temperature correction
using a temperature sensor. If possible the calibration should be carried out
at the same temperature as that at which the later measurements will be
carried out.

pH

14

7

0

100

200

300

400

500

-500

-400

-300

-200

-100

U[mV]

d U

d pH

*

pH

as

Slope =

dU/dpH

U

N

(T)

Temperature 25°C

Without calibration the relative slope of the electrode system will be 1
and pH

as

will be set at 7.

Depending on the requirements for accuracy the calibration can be carried
out with 1, 2 or up to 9 buffer solutions.

Advertising