12.1 Error messages

796 Titroprocessor, Instructions for use

242

242

FILE

FILE 1:

wrong file type "XXXXXXXX"

The selected file does not correspond to the type of file required.

FILE 2:

not enough memory "XXXXXXXX"

The selected memory area is full. Delete files which are no longer required
and try again to save.

FILE 3:

data inconsistent "XXXXXXXX"

The determination file has been damaged. Delete it.

FILE 5:

verification of checksum failed "XXXXXXXX"

The file mentioned is damaged. Delete it.

FILE 6:

incompatible version of "XXXXXXXX"

The method mentioned was created with a different program version of the
796 Titroprocessor and cannot be used on this instrument.

FILE 7:

checksum failed "XXXXXXXX"

The file mentioned is damaged. Delete it.

FILE 8:

can not delete current det. "XXXXXXXX"

An attempt has been made to delete the current actual determination from
the working memory. If as much working memory space as possible must
be made available for a series of samples then initialize the statistics func-
tion. On the main page enter 'ini'=initialize under statistics.

FORMULA

FORMULA 1:

formula list is full

No more formulas can be entered in the table.

FORMULA 2...11 :formula X:

Syntax error in the calculation formula. Correct the formula.

FORMULA 24...30:XXXXXXXX:

Invalid values have occurred during the calculation of a formula. The result
is incorrect. Check the corresponding calculation formula and try using the
TST() function to replace a possibly invalid value in the calculation with a
different value, see page 122ff.