Specifications

Mains (line):

115/230 V~50/60 Hz, 175 VA


Fuse:

2 A, time-delay, 5 x 20 mm


Pressure:

6.9 kPa (1 psi) operating, 6.9 MPa (1000 psi) max.

working


Maximum temperature
PD-D2-IS:

400°C

PD-D2:

400°C


Heater power:

60 W max., 48 V, PWM


Sensitivity

10X

range:

1.0

V/nA

1X range:

0.1 V/nA ± 1%*

Range

10X:

10 nA full-scale

1X:

100 nA full-scale*


Noise

1X:

20

fA/√Hz (referred to the input), 0.1 - 10 Hz*

10X:

5 fA/√Hz (referred to the input), 0.1 - 10 Hz


Risetime:

10 msec, 10% = 90%*


Output impedance: < 1 Ω,

attenuated

1

KΩ, unattenuated, 1 V, 10 V

Environmental ratings: 10°C - 50°C, 10% - 95% relataive humidity

*Controller only, PDHID mode, 10 V unattenuated output