System Requirements

Gas Specifications

For Model D-2

Detector Mode

PDECD

PDHID

Ar-PDPID

Kr-PDPID

Xe-PDPID

Discharge gas

Helium

Helium

2% Ar in He 1.5% Kr in He 0.8% Xe in He

Carrier gas

Helium

Helium

*

**

**

Dopant gas

3% Xe in He

n/a

n/a

n/a

n/a

Detector Mode

PDHID

Ar-PDPID

Kr-PDPID

Xe-PDPID

Discharge gas

Helium

2% Ar in He 1.5% Kr in He 0.8% Xe in He

Carrier gas

Helium

*

**

**

*

*

Any gas including He which has an ionization potential greater than 12 eV

**

Any gas including He which has an ionization potential greater than 11 eV

The helium purifier (HP2) supplied with the system must be used for the discharge gas.

Purity Specifications

Helium (discharge and carrier gas) must have a minimum purity of
99.999%, with < 20 ppm Ne impurity. For trace analysis of fixed gases,
we strongly recomment 99.9999% purity He with < 0.5 ppm Ne.

Ar-PDPID mode:

2% ± 0.3% Ar in 99.999% He balance

Kr-PDPID mode:

1.5% ± 0.1% Kr in 99.999% He balance

Xe-PDPID mode:

0.8% ± 0.2% Xe in 99.999% He balance

For Model D-2-I

*

*

Any gas including He which has an ionization potential greater than 12 eV

**

Any gas including He which has an ionization potential greater than 11 eV

Purity Specifications

Helium (discharge and carrier gas) must have a minimum purity of
99.999%, with < 20 ppm Ne impurity. For trace analysis of fixed gases,
we strongly recomment 99.9999% purity He with < 0.5 ppm Ne.

Ar-PDPID mode:

2% ± 0.2% Ar in 99.999% He balance

Kr-PDPID mode:

1.5% ± 0.1% Kr in 99.999% He balance

Xe-PDPID mode:

0.8% ± 0.2% Xe in 99.999% He balance