Gas specifications, For model d-2, For model d-2-i – VICI D-2-I User Manual

Page 15

Advertising
background image

11

System Requirements

Gas Specifications

For Model D-2

Detector Mode

PDECD

PDHID

Ar-PDPID

Kr-PDPID

Xe-PDPID

Discharge gas

Helium

Helium

2% Ar in He 1.5% Kr in He 0.8% Xe in He

Carrier gas

Helium

Helium

*

**

**

Dopant gas

3% Xe in He

n/a

n/a

n/a

n/a

Detector Mode

PDHID

Ar-PDPID

Kr-PDPID

Xe-PDPID

Discharge gas

Helium

2% Ar in He 1.5% Kr in He 0.8% Xe in He

Carrier gas

Helium

*

**

**

*

*

Any gas including He which has an ionization potential greater than 12 eV

**

Any gas including He which has an ionization potential greater than 11 eV

The helium purifier (HP2) supplied with the system must be used for the discharge gas.

Purity Specifications

•

Helium (discharge and carrier gas) must have a minimum purity of
99.999%, with < 20 ppm Ne impurity. For trace analysis of fixed gases,
we strongly recomment 99.9999% purity He with < 0.5 ppm Ne.

•

Ar-PDPID mode:

2% ± 0.3% Ar in 99.999% He balance

•

Kr-PDPID mode:

1.5% ± 0.1% Kr in 99.999% He balance

•

Xe-PDPID mode:

0.8% ± 0.2% Xe in 99.999% He balance

For Model D-2-I

*

*

Any gas including He which has an ionization potential greater than 12 eV

**

Any gas including He which has an ionization potential greater than 11 eV

Purity Specifications

•

Helium (discharge and carrier gas) must have a minimum purity of
99.999%, with < 20 ppm Ne impurity. For trace analysis of fixed gases,
we strongly recomment 99.9999% purity He with < 0.5 ppm Ne.

•

Ar-PDPID mode:

2% ± 0.2% Ar in 99.999% He balance

•

Kr-PDPID mode:

1.5% ± 0.1% Kr in 99.999% He balance

•

Xe-PDPID mode:

0.8% ± 0.2% Xe in 99.999% He balance