Electron Capture Mode (PD-D2 only)

Since the PDHID mode provides a better indication of the cleanliness and
the integrity (leak-tightness) of the system, we utilize that mode for initial
testing. If the system is operating according to the parameters described
thus far, it is ready for operation in the PDHID mode. To configure the
PD-D2 to operate in the PDECD mode:

1. If the discharge is not already established, turn on the

MAINS switch and

the

DISCHARGE switch. The discharge should start within five minutes.

2. Set the

MODE switch on the controller to PDECD.

3. Set the detector temperature with the

TEMPERATURE control knob.

4. Set the electrometer

RANGE switch to 1X. (The high detector back-

ground current associated with the PDECD mode normally precludes
operation in the 10X range).

5. Leave the

BIAS-B cable alone. Remove the BIAS-A cable and move

the

ELECTROMETER cable to the bottom electrode. Remember that

connections should be finger-tight only.

High voltage will be present on the BIAS-A cable
during operation. Cover the end of the cable with
tape or otherwise protect the end from accidental
contact with persons or equipment.

6. Disconnect the column and leave the column inlet open. Reinstall the

column as indicated in on page 17. The suggested range for the column
position is from 99 to 102 mm, wherein the detector response (not the
background current) is within 90% of the maximum. Since there may be
some variation in the flow rate for the different types of capillary columns,
you may need to optimize the column position within this suggested
range.

7. With the dopant line still disconnected, set the dopant gas (3% xenon in

helium) flow rate to 5 - 7 mL/min (±0.2 mL/min). (In some circumstances,
methane in helium may be used as the dopant to affect a cost savings.
Consult the factory for recommendations.)

8. Remove the plug from the DOPANT port on the detector and replace it

with the dopant gas line.

9. Check the

BIAS-B voltage, indicated by the controller DISPLAY. If the

displayed voltage is 0.5 - 1 V at room temperature, the system is ready
for operation. (A bias voltage less than 0.5 V can result in an unstable,
erratic, and noisy system.)

In the PDECD mode, the bias voltage is internally controlled to obtain a

constant detector current, set by means of the

CURRENT switch. The

power-up default value for this current is 10 nA at 1X (1 nA at 10X), which
does not normally need to be changed. A higher set current results in a
more sensitive detector, but with a correspondingly reduced linear range.