VICI D-2-I User Manual

Page 20

Advertising
background image

16

Installing and Purging the Helium Purifier

1. If the pressure regulator has a 1/8" male cone-type outlet port, install

the Valco 1/8" external to 1/16" internal reducer (EZR21); if it has a 1/4"
male cone-type outlet port, install the Valco 1/4" external to 1/16" internal
reducer (EZR41). For other regulator outlet fittings, a wide variety of
Valco adapters are available.

2. Remove the cap from the inlet tube of the Valco helium purifier and insert

the tube fitting into the 1/16" reducer port. (Keep the outlet tube capped.)
Use a 1/4" wrench to turn the nut one-quarter turn past the point where
the ferrule first starts to grab the tubing. Do not remove the fitting. When
made up properly, it should be leak-tight.

3. Turn the output pressure regulating knob clockwise until the gauge

registers 345 KPA (50 psi).

4. Allow five minutes for equilibration, then turn the regulating knob all the

way counterclockwise.

5. Observe the needle of the output pressure gauge for 15 minutes. There

will be a slight initial drop, but if it doesn’t move after that, consider that
all the connections are tight.

6. If necessary, use an electronic leak detector to locate any leaks. If a

leak detector is not available, tighten all the fittings (including the output
pressure guage), and repressurize the system for another test.

CAUTION: Never use leak detecting fluids on any part of
this system.

7. Upcap the outlet tube of the purifier and purge the system for 15 to 30

minutes at 60 - 80 mL/min to eliminate air from the purifier getter material.

Connecting the Discharge Gas to the Detector

1. If you are supplying the GC from the helium purifier, use the Valco tee

(ZT1). Otherwise, use one of the Valco 1/16" unions (ZU1) to connect
the outlet tube of the purifier to the inlet of the supplied discharge gas
restrictor (TGA-R-30F60P or TGA-R-50F50P).

2. Connect the outlet end of the restrictor to a flow measuring device and

adjust the helium pressure to obtain a flow of ~30 mL/min for the PD-D2
and ~50 mL/min for the PD-D2-IS.

3. After setting the flow rate, connect the outlet of the restrictor to the

discharge gas inlet tube at the top of the detector.

Installation

EZR21