Installation

Electrical Connections

CAUTION: Do not use a wrench to tighten the SMC
connectors on the bias and electrometer cables.
Connections should be finger tight only.

1. Referring to

Figure 6 as necessary, connect the bias and electrometer

cables as follows: the

BIAS-B cable to the top electrode (closest to the

discharge electrodes), the electrometer cable (

ELECT) to the middle,

and the

BIAS-A cable to the bottom electrode. This setup is the same

for both detector models.

2. Connect the high-voltage cable from the detector to the pulse supply, and

connect the pulse supply cable between the back of the controller and
the pulse supply. Connect the heater cable from the detector to the back
of the controller.

CONTROLLER

(Rear panel)

HEATER CABLE

FROM DETECTOR

DISCHARGE

MODULE

HIGH

VOLTAGE

ECD MODE: BIAS-B

OPEN

ELECT.

HID MODE: BIAS-B

ELECT.
BIAS-A

ELECTROMETER

BIAS-A

BIAS-B

RED
BLACK
WHITE
GREEN
BROWN
BLUE
ORANGE
YELLOW

RED
BLACK
GREEN
WHITE

0 - 10 V
Common
0 - 1 V
Shield
COM
SET

UNATTENUATED

OUTPUT

AUTO

ZERO

COMPUTER OR

INTEGRATOR

+

ATTENUATED

OUTPUT

SHIELD

RECORDER

DETECTOR

GROUND

Figure 6: Electrical connections

If your controller rear
panel does not look like
this, refer to page 31 in
the Appendix.