COMPUTER CONNECTION

PC LINK

The 2020we/wi may be interfaced with any Windows-based computer by

using the LaMotte SMARTLink 3 Program and USB Cable. The program

will store test information and results in a database. To transfer data from

the meter to a computer, plug the smaller end of the USB cable (USB

mini B connector) into the meter and the larger end of the USB cable

(USB Type A connector) into a USB port on a computer. The 2020we/

wi will send the following data: test name, wavelength, concentration,

transmittance, absorbance, sample, blank, time of test, and date of test.

OUTPUT

USB

COMPUTER CONNECTION

USB Type A, USB mini B, Order Cable Code 1720.

BATTERY/AC OPERATION

The 2020we/wi may be operated on battery power or using USB wall

adapter or USB computer connection. If using the meter as a bench top

unit, use the wall adapter if possible to extend the battery life. The meter

will remain on when the USB adapter is used.
To charge the battery with the wall adapter, plug the smaller end of the

USB cable (USB mini B connector) into the meter and the larger end of

the USB cable (USB Type A connector) into the wall adapter. Plug the

wall adapter into an AC outlet. Reinsert the USB port plug after charging.
To charge the battery from a computer, plug the smaller end of the USB

cable (USB mini B connector) into the meter and the larger end of the

USB cable (USB Type A connector) into a USB port on a computer.

Reinsert the USB port plug after charging.
The battery icon will show no bars and fl ash when the unit fi rst turns on.

Then the indicator will indicate the battery status by showing 0, 1, 2, 3 or

4 bars.
It will take 5 hours to fully charge a low battery. The battery icon will

fl ash when the battery is charging. The battery icon will show four bars

and stop fl ashing when it is fully charged. The charging circuit will

automatically switch to a fl oat charge when the battery is fully charged.

The charger may remain connected. Some computers will NOT supply

power to their USB ports during standby operation. The wall adapter will

charge the unit continuously.
The battery icon will show no bars and continuously fl ash if the battery

is getting low but the unit will still operate normally. A “Low Battery”

message on the status bar of the display will replace the time when the

battery voltage is too low for proper operation and accuracy may be

degraded. A “Shutdown Low Batt” message on the display will appear

for a few seconds before the power is switched off when the battery is

too low to operate the unit.