background image

4. Press

ENTER

to select

Turbidity - With Blank.

Turbidity WB (F)


Scan Blank

Scan Sample

12:00:00 001/500

5. Rinse a clean tube (0290)

three times with the blank.

If samples are expected to

read below 1 NTU the meter

should be blanked with a 0

NTU Primary Standard or

prepared turbidity-free (<0.1

NTU) water. For the most

accurate results, use the

same tube for the blank and

the sample.

l

6. Fill the tube to the fi ll line with

the blank. Pour the blank

down the inside of the tube to

avoid creating bubbles. Cap

the tube.

7. Wipe the tube thoroughly with

a lint-free cloth.