LaMotte 2020wi Portable Turbidity Meters User Manual

Page 16

Advertising
background image

Acronyms Defi nitions

Notes

Regulatory

Method

NTU

Nephelometric

Turbidity Units

Incandescent white

light between 400 and

600 nm, 90° detection,

2020we

EPA 180.1

FNU

Formazin

Nephelometric

Units

IR LED (usually) 860

nm, bandwidth less than

60 nm, 90° detection,

2020wi

ISO 7027

AU

Attenuation

Units

Incandescent white

light between 400 and

600 nm, 180° detection,

2020we

Not

applicable

FAU

Formazin

Attenuation

Units

IR LED (usually) 860

nm, bandwidth less than

60 nm, 180° detection,

2020wi

ISO 7027

ASBC

American

Society of

Brewing

Chemists

2020we/wi

Not

applicable

EBC

European

Brewery

Convention

2020we/wi

Not

applicable

The 2020we and 2020wi each use two turbidity units. The 2020we

reports the result as NTU when the 90° detector is used and AU when

the 180° detector is used. The 2020wi reports the result as FNU when

the 90° detector is used and FAU when the 180° detector is used.

TAKING TURBIDITY WATER SAMPLES

Clean plastic or glass containers may be used for turbidity samples.

Ideally, samples should be tested soon after collection and at the same

temperature as when collected.

16