LaMotte 2020wi Portable Turbidity Meters User Manual

Page 18

Advertising
background image

AVERAGING

The averaging option allows the user to average multiple readings. This

option will improve the accuracy of samples with readings that may

tend to drift with time. When the two, fi ve or ten measurement option

has been selected the fi nal average is displayed. The default setting is

disabled. To change the setting:

1. Press and briefl y hold

to turn the meter on. The

LaMotte logo screen will

appear for about 3 seconds

and the

Main Menu will

appear.

Main Menu

Measure

Data Logging
Options
Run PC Link

12:00:00 001/500

2. Press

to scroll to

Options.

Main Menu

Measure
Data Logging

Options

Run PC Link

12:00:00 001/500

3. Press

ENTER

to select

Options.

Options Menu

Averaging

Turbidity Options
Set Clock
Set PWR Save

12:00:00 001/500

4. Press

ENTER

to select

Averaging.

Averaging Menu

Disabled

2 Measurements
5 Measurements
10 Measurements

12:00:00 001/500

18

Options/Set Up