retain the old calibration data for the other ranges. It is recommended

that the meter be calibrated for each range that will be used. The value

of the standards chosen for the calibration should not be at the extremes

of the ranges (11, 110, 300, 600 NTU/FTU). The meter is auto-ranging

and will automatically select the appropriate range for the sample being

tested.
It is recommended that the meter be calibrated daily.

1. Press and briefl y hold

to turn the meter on. The

LaMotte logo screen will

appear for about 3 seconds

and the

Main Menu will

appear.

Main Menu

Measure

Data Logging
Options
Run PC Link

12:00:00 001/500

2. Press

ENTER

to select

Measure.

Measure Menu

Turbidity - No Blank

Turbidity - With Blank

12:00:00 001/500

3. Press

to scroll to

Turbidity - With Blank.

Measure Menu

Turbidity - No Blank

Turbidity - With Blank

12:00:00 001/500

4. Press

ENTER

to select

Turbidity - With Blank.

Turbidity WB (F)

Scan Blank

Scan Sample

12:00:00 001/500

37