SAMPLE DILUTION TECHNIQUES

If a test result is out of the range of the meter, it must be diluted. The

test should then be repeated on the diluted sample. The following table

gives quick reference guidelines for dilutions of various proportions.

Amount of Sample Deionized Water to Bring

Final Volume to 10 mL

Multiplication Factor

10 mL

0 mL

1

5 mL

5 mL

2

2.5 mL

7.5 mL

4

1 mL

9 mL

10

0.5 mL

9.5 mL

20

All dilutions are based on a fi nal volume of 10 mL, so several dilutions

will require small volumes of the water sample. Graduated pipets should

be used for all dilutions. If volumetric glassware is not available, dilutions

can be made with the colorimeter tube. Fill the tube to the 10 mL line

with the sample and then transfer it to another container. Add 10 mL

volumes of deionized water to the container and mix. Transfer 10 mL of

the diluted sample to the colorimeter tube and follow the test procedure.

Repeat the dilution and testing procedures until the result falls within

the range of the calibration. Multiply the test result by the dilution factor.

For example, if 10 mL of the sample water is diluted with three 10 mL

volumes of deionized water, the dilution factor is four. The test result of

the diluted sample should be multiplied by four.

OPTIONS & SET UP

FACTORY DEFAULT SETTINGS

Settings that have user options have been set at the factory to default

settings.
The factory default settings are:

Averaging

Disabled

Turbidity Units

NTU

Turbidity Calibration

Formazin

Date Format

MM-DD-YYYY

Power Save

5 minutes

Backlight

10 seconds

Language

English