LaMotte 2020wi Portable Turbidity Meters User Manual

Page 46

Advertising
background image

13. Open the meter lid. Insert the

tube into the chamber. Align

the index line on the tube with

the index arrow on the meter.

Close the lid.

14. Press

ENTER

to scan the

standard. The screen will

display

Reading for about 1

second. The result will appear

on the screen.

Turbidity WB (F)

0.99

NTU

Scan Blank

Scan Sample

12:00:00 001/500

NOTE: The meter will remember the last scanned blank reading. It is not

necessary to scan a blank each time the test is performed. To use the

previous blank reading, instead of scanning a new one, scroll to Scan

Sample and proceed. For the most accurate results, the meter should

be blanked before each test and the same tube should be used for the

blank and the reacted sample.

DILUTION PROCEDURES

If a sample is encountered that is more than 4000 NTU or FNU, a careful

dilution with 0 NTU/FNU or very low turbidity water will bring the sample

into an acceptable range. However, there is no guarantee that halving

the concentration will exactly halve the NTU or FNU value. Particulates

often react in an unpredictable manner when diluted.
Turbidity-Free Water

The defi nition of low turbidity and turbidity-free water has changed as

fi lter technology has changed and nephelometric instruments have

become more sensitive. At one time turbidity-free water was defi ned as

water that had passed through a 0.6 micron fi lter. Now 0.1 micron fi lters

are available and higher purity water is possible. Water that has been

passed through a 0.1 micron fi lter could be considered particle free and

therefore turbidity free, 0 NTU water. Turbidity is caused by scattered

light. Therefore, low turbidity water is water without any particles that

scatter a measurable amount of light. But water that passed through

a 0.1 micron fi lter may still have detectable light scatter with modern

instruments. This light scattering can be the result of dissolved

molecules or sub-micron sized particles that can not be fi ltered out of

the water. Because there may still be a small amount of scattered light

from dissolved molecules, high purity water is often called low turbidity

water and assigned a value of 0.01 or 0.02 NTU. However, because this

water is used as a baseline to compare to sample water, the difference

between the sample and the low turbidity or turbidity-free water will

46

Analysis

Advertising
This manual is related to the following products: