PROCEDURE
1.
Press and hold
ON
button until colorimeter turns on.
2.
Press
ENTER
to start.
3.
Press
ENTER
to select TESTING MENU.
4.
Select ALL TESTS (or another sequence containing 29 Copper
BCA-LR
) from TESTING MENU.
5.
Scroll to and select 29 Copper BCA-LR from menu.
6.
Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with
sample.
7.
Insert tube into chamber, close lid and select SCAN BLANK.
8.
Remove tube from colorimeter and add one *Copper Tablet (T-3808).
Cap and shake vigorously until tablet dissolves. Solution will turn purple if
copper is present. Wait 2 minutes.
9.
At end of 2 minute waiting period, mix, insert tube into chamber, close lid
and select SCAN SAMPLE. Record result.
10.
Press
OFF button to turn colorimeter off or press EXIT button to exit to a
previous menu or make another menu selection.
Copper–LR 2/2