Storing – LaMotte SMART 2 User Manual
Page 32

A reverse font (dark background with light
characters) will appear to indicate that the
reading can be adjusted. Use
s or t to scroll to
the concentration of the sample, 0.10 ppm in this
example.
002 Alumninum
0.09
* Calibrate
Set the calibration by pressing
Q/ENTER to
select Calibrate.
002 Aluminum
0.10
* Calibrate
Two menu choices will be offered, Set
Calibration
and Factory Setting. Set
the calibration by pressing
Q/ENTER to select
Set Calibration
; or use
s or t to scroll to
and select Factory Setting t o revert to the
factory calibration.
002 Aluminum
0.10
* Set Calibration
Factory Setting
The meter will display the message “Storing”
and return to 002 Aluminum test.
Storing
The calibration for 002 Aluminum has now
been standardized and can be used for testing.
The standardization can be removed by repeating
the calibration and selecting Factory
Setting
.
002 Aluminum
* Scan Sample
Next Test
Previous Test
%/Abs
Calibrate
Scan Blank
32
SMART2 COLORIMETER 1.07