Blank done – LaMotte SMART 2 User Manual
Page 34

Insert blank, press the
Q/ENTER button to scan
blank.
102 Abs 520
* Scan Blank
The screen will display Blank Done for about
1 second.
102 Abs 520
Blank Done
* Scan Blank
Insert the reacted sample. Press the
Q/ENTER
button to scan the sample.
102 Abs 520
* Scan Sample
Record test result. To repeat the test, press the
Q/ENTER button to scan the sample again. The
last blank scanned is used to zero the colorimeter
for repeated scans. A different blank can be used
by pressing the
s button to scroll back to Scan
Blank
and then scanning another blank. Scroll
with
t or s and make another selection with
Q/ENTER. The %T or Absorbance of the last
test can be viewed by choosing %T/Abs. Press
EXIT to escape to previous menus.
NOTE: The menus loop in this screen so either
t or s will lead to the menu selection needed.
NOTE: The Calibrate function does not work in
the Absorbance mode.
102 Abs 520
0.95
* Scan Sample
Next Test
Previous Test
%T/Abs
Calibrate
Scan Blank
34
SMART2 COLORIMETER 1.07