PROCEDURE

1.

Press and hold ON button until colorimeter turns on.

2.

Press

ENTER to start.

3.

Press

ENTER to select TESTING MENU.

4.

Select ALL TESTS (or another sequence containing 69 Nitrite-TT)
from TESTING MENU.

5.

Scroll to and select 69 Nitrite-TT from menu.

6.

Rinse a tube (0290) with sample water. Fill to 10 mL line with sample.

7.

Insert the tube into chamber, close lid and select SCAN BLANK.

8.

Remove the tube from colorimeter.

9.

Add one *Nitrite Spectrophotometric Grade Tablet (3886).

10.

Use Tablet Crusher (0175) to crush tablet.

11.

Cap tube.

12.

Shake vigorously for 20 seconds.

13.

Wait 2 minutes.

14.

Immediately, insert tube into chamber, close lid and select SCAN
SAMPLE

. Record result in ppm nitrite.

15.

Press

OFF button to turn the colorimeter off or press EXIT button to exit

to a previous menu or make another menu selection.

þ NOTE: For best possible results, a reagent blank should be determined to

account for any contribution to the test result by the reagent system. To
determine the reagent blank, follow the above test procedure to scan a distilled
or deionized water blank. Then follow the above procedure to perform the test
on a distilled or deionized water sample. This test result is the reagent blank.
Subtract the reagent blank from all subsequent test results of unknown samples.

To convert nitrite (NO

2

) results to nitrite-nitrogen (NO

2

–N), divide results by

3.3.

Nitrite-TesTab 2/2

Smart2 TEST PROCEDURES 1.07