LaMotte WaterLink 3 Express Lab User Manual

Page 11

Advertising
background image

sample without inserting the tip much beyond the foil opening. If reagent

does get on the tip, rinse off the tip before proceeding to the next test.

TIP: When water sample is dispensed too rapidly, air bubbles could

develop and result in high results. Try inverting the vial again or tapping it

on a firm surface to dislodge the bubbles.

Continue filling the UDVs from right to left, finishing with Free Chlorine.

8.

The vials must be filled with the pool or spa water sample from right

to left and read from left to right to allow for proper color development

time.
Press

9.

ENTER

to select BLANK Vial. Press

ENTER

again to read blank.

BLANK DONE will appear on the display. The colorimeter is now ready

to test for Free Chlorine.

NOTE: The lid does not need to be closed when testing with UDVs.
Remove the Blank vial.

10.

Invert the Free Chlorine UDV 3 times to mix. Finger protectors are

11.

included to avoid stains and to protect individuals with sensitive skin

while mixing the UDVs.

NOTE: If clumps of powder are lodged in the corner of the UDV, tap the

vial on the counter then invert again. Should water get on the outside of

the UDV, wipe the UDV with a dry cloth before inserting it into the meter

chamber.

Insert the Free Chlorine UDV into the colorimeter chamber. Press

ENTER

to take a measurement and proceed to the next test. Record result.

NOTE: When using the colorimeter with the DataMate computer

program, the UDV must remain in the chamber until after the button is

pushed and the test result appears on the colorimeter screen. Once the

next test name appears the UDV can be removed and replaced with the

next UDV and

ENTER

can be pressed. This process should be done until

all of the tests have been performed.

NOTE: To BACK UP while in the middle of a test series press

or

which will scroll through the Test Sequence.

Remove the Free Chlorine UDV. Continue testing from left to right,

12.

inverting each vial 3 times before placing it into the meter. Finish testing

on the right with the Borate UDV. Press

EXIT

to go back to the Main

Menu or

to turn meter off.

NOTE: Individual UDV Tests

The WaterLink

®

3 Express meter is designed to provide accurate results

when all of the UDV tests are run in sequence. To run an individual test for

a selected test factor, wait the specified time listed on the counter mat after

adding the sample before mixing and measuring the test result. Invert the

UDV 3 times then place it in the meter chamber. If clumps of powder are

lodged in the corner, tap the vial on the counter then invert.

11

Testing with the W

aterLink

®

3 Express Lab

Advertising