BIGUANIDE SHOCK PROCEDURE
Remove the UDV Adapter (1724) from the meter chamber. (See page 8).

Press and briefly hold

1.

to turn the meter on.

Press

2.

to scroll to Other Series.

Press

3.

ENTER

to select Other Series.

Press

4.

to scroll to Biguanide Shock.

Press

5.

ENTER

to select Biguanide Shock.

Rinse a clean, glass test tube (0290) with sample water. Fill to the 10

6.

mL line with the sample.
Insert the tube into the chamber. Press

7.

ENTER

to select BLANK Vial.

NOTE: If a blank has been scanned previously in the same type of

tube, it is not necessary to scan a blank again.
Remove the tube from colorimeter and add 4 drops of *Hydrogen

8.

Peroxide Reagent #1 (6452). Cap and mix.
Add one *Hydrogen Peroxide LR Tablet (6454A). Crush tablet with

9.

Tablet Crusher (0175). Cap and mix for 30 seconds. Solution will turn

pink if hydrogen peroxide is present. Wait 5 minutes for full color

development.
At the end of 5 minute waiting period, mix, insert tube into chamber,

10.

close lid and select SCAN SAMPLE. Record result as ppm Biguanide

Shock.
Follow the next test procedure to test the next test factor in the series,

11.

press

or

to select another test in the series, press

EXIT

to

escape to a previous menu, or press

to turn the meter off.