LaMotte WaterLink 3 Express Lab User Manual

Page 16

Advertising
background image

BIGUANIDE PROCEDURE
Remove the UDV Adapter (1724) from the meter chamber. (See page 8)

Press and briefly hold

1.

to turn the meter on.

Press

2.

to scroll to Other Series.

Press

3.

ENTER

to select Other Series.

Press

4.

to scroll to Biguanide.

Press

5.

ENTER

to select Biguanide.

Rinse a clean, glass test tube (0290) with sample water. Fill to the 10

6.

mL line with the sample.
Insert the tube into the chamber. Press

7.

ENTER

to select BLANK Vial.

NOTE: If a blank has been scanned previously in the same type of

tube, it is not necessary to scan a blank again.
Remove the tube from colorimeter.

8.

Use the 1.0 mL pipet (0354) to add 2.0 mL of Biguanide Indicator

9.

(3994). Cap and invert three times to mix.
Wait 1 minute.

10.

Insert the tube into chamber. Close lid.

11.

Select

12.

SCAN SAMPLE. Record result as ppm Biguanide

Follow the next test procedure to test the next test factor in the series,

13.

press

or

to select another test in the series, press

EXIT

to

escape to a previous menu, or press

to turn the meter off.

16

Testing with the W

aterLink

®

3 Express Lab