MANGANESE PROCEDURE
Remove UDV Adapter (1724) from the meter chamber. (See page 8)

Press and briefly hold

1.

to turn the meter on.

Press

2.

to scroll to Other Series.

Press

3.

ENTER

to select Other Series.

Press

4.

to scroll to Manganese.

Press

5.

ENTER

to select Manganese.

Rinse a clean, glass test tube (0290) with sample water. Fill to the 10

6.

mL line with the sample.
Insert the tube into the chamber. Press

7.

ENTER

to select BLANK Vial.

NOTE: If a blank has been scanned previously in the same type of

tube, it is not necessary to scan a blank again.
Remove tube from colorimeter. Use the 1.0 mL pipet (0354) to add 2.0

8.

mL (two measures) of *Hardness Buffer Reagent (4255). Swirl to mix.
Add 2 drops of *Sodium Cyanide, 10% (6565). Cap and mix.

9.

Use the 0.5 mL pipet (0369) to add 0.5 mL of *Manganese Indicator

10.

Reagent (3956). Cap and mix.
Immediately insert tube into chamber, close lid and select

11.

SCAN

SAMPLE. Record result as ppm Manganese.
Follow the next test procedure to test the next test factor in the series,

12.

press

or

to select another test in the series, press

EXIT

to

escape to a previous menu, or press

to turn the meter off.