CFX96 Touch, CFX96 Touch Deep Well, CFX Connect, and CFX384 Touch Systems Instruction Manual

97

The End Point tab shows the average RFU values to determine whether or not the target was
amplified by the last (end) cycle. Use these data to determine if a specific target sequence is
present (positive) in a sample. Positive targets have higher RFU values than the cutoff level you
define.

TIP: To create an end point protocol, open the Protocol tab (Run Setup window)
and select Options > End Point Only Run.

The software displays these data in the End Point tab:

Settings. Adjust data analysis settings

Results. Shows the results immediately after you adjust the Settings

Well Selector. Select the wells with the end point data you want to show

Well spreadsheet. Shows a spreadsheet of the end RFU collected in the selected wells

Figure 73. Layout of the End Point analysis tab.

The Results list includes this information:

Lowest RFU value. Lowest RFU value in the data

Highest RFU value. Highest RFU value in the data

Negative Control Average. Average RFU for the wells that contain negative controls

Cut Off Value. Calculated by adding the tolerance (RFU or Percentage of Range listed in
the Settings) and the average of the negative controls. Samples with RFUs that are
greater than the cutoff value will be called “Positive.” To adjust the cutoff value, change
the RFU or Percentage of Range

The Cut Off Value is calculated using this formula:

Select a tolerance by one of these methods:

RFUs (default). Select this method to use an absolute RFU value for the tolerance. The
minimum RFU tolerance value is 2. The maximum is the absolute value of the highest
RFU value minus the absolute value of the lowest RFU value. The default RFU tolerance
value is 10% of the total RFU range

Percent of Range. Select this method to use a percentage of the RFU range for the
tolerance. The minimum percent of range is 1%. The maximum percent of range is 99%.
The default percent of range is 10%

Cut Off Value

Negative Control Average

Tolerance

+

=