Bio-Rad Firmware & Software Updates User Manual
Page 170

CFX96 Touch, CFX96 Touch Deep Well, CFX Connect, and CFX384 Touch Systems Instruction Manual
157
the protocol continues only if the software is not shut down by the computer. When
the computer and software start up again, the protocol continues
If you want to open a locked motorized lid on a reaction module to remove your samples
during a power failure, follow these steps to remove the locking plate:
1. Remove the reaction module from the C1000 Touch or CFX Connect chassis by pushing
down on the locking bar.
2. Position the module on the front of a desk so that the front of the module extends 2
inches over the edge of the desk as shown in Figure 107.
Figure 107. Setting up the Optical Module to remove the locking plate.
3. With an allen wrench, remove the two large screws from under the front edge of the
reaction module (below the button for opening the lid). Do not remove the two small
screws located at the front edge of the module. You should hear the locking latch release
from inside the module. Figure 108 shows the two large screws.
Figure 108. Remove these screws to unlock the optical module.
4. Push the reaction module lid open. Notice that the latch (dark plastic) is no longer
attached. Remove your samples from the block.