Bio-Rad Model 583 Gel Drying Supports User Manual

Page 12

Advertising
background image

9

Gel Preparation Techniques

Table 4 (continued)

Drying Gels

Step

Procedure

6.

Program the drying cycle.

Description

For Sequencing gels:

a.

Select the

Sequencing

cycle.

b.

For normal rapid drying, the default temperature is
80°C.

For gels containing fluors for autoradiography, set
the temperature to 60°C.

c.

Select a drying time. Refer to Chapter 4, Drying
Times, for recommended drying times.

For gels having < 14% constant concentration acry-
lamide:

a.

Select the

PAGE

cycle.

b.

For normal rapid drying, set a temperature of
80°C.

For gels containing fluors for autoradiography,
select 60°C.

c.

Select a drying time. Refer to Chapter 4 for rec-
ommended drying times.

For gradient gels, gels having > 15% constant concen-
tration acrylamide, and 1.5 mm gels:

a.

Begin by soaking the gel for about one hour in a
500 ml solution of 3% glycerol (5%, if the gradient
is very steep), 40% MeOH, and 10% HOAc.
Gently agitate during the entire soak period.

b.

Select the

GRADIENT

cycle.

c.

Set a temperature of 60°C to 70°C.

d.

Select a drying time. Refer to Chapter 4 for rec-
ommended drying times.

Advertising